Getting ready for WEL: Updating the CO TWA alarm on your Altair 4X

Getting ready for WEL: How to change the CO TWA alarm on your MSA Altair 4X


Part of our Getting Ready for WEL series: practical steps to prepare your workplace for Australia’s new Workplace Exposure Limits.

On 1 December 2026, Australia replaces the Workplace Exposure Standards (WES) with the new Workplace Exposure Limits (WEL). Many limits are changing, and one that affects almost every site using portable gas detectors is carbon monoxide.

The CO 8-hour time-weighted average (TWA) limit drops from 30 ppm to 20 ppm.

If your detectors are still set to the old limit, or are running factory defaults that were never set for Australian limits, your TWA alarm won’t go off until a worker is already over the new limit.

The good news: on an MSA Altair 4X, updating it takes about two minutes per unit.

What’s changing and when

Until 30 November 2026 (WES) From 1 December 2026 (WEL)
Carbon monoxide, 8-hour TWA 30 ppm 20 ppm

The TWA is the average concentration a worker can be exposed to over an 8-hour working day. Your detector’s TWA alarm tracks that running exposure across the shift, so it needs to match the limit you’re managing to.

Does this apply in Victoria?

Yes. Victoria is adopting the national WEL list from 1 December 2026. There is a three-year transition (to 1 December 2029) for 15 specific substances, such as bitumen fumes, diesel particulate matter, grain dust, styrene and sulphur dioxide. Carbon monoxide is not one of them, so the 20 ppm limit applies in Victoria from 1 December 2026.

WorkSafe Victoria has also made it clear that a transition period isn’t a reason to wait. Employers are expected to start acting now.

Before you start

  • Have the detector in clean air, with a charged battery.
  • Have the setup password ready. The factory default is 672, unless your organisation has changed it.
  • Check your company’s gas detection procedure. Alarm setpoints should be a controlled decision, documented and applied consistently across your fleet, not something individuals adjust on their own.

Step by step: changing the CO TWA alarm

  1. Enter setup mode. With the unit off, press and hold the ▲ button, then press ON/OFF. Setup mode can only be entered at power-on.
  2. Enter the password. Use ▲ / ▼ to enter 672, then press ON/OFF. The unit beeps three times when the password is accepted.
  3. Skip to Alarm Setup. The menus appear in this order: Sensor Setup, Calibration Setup, Alarm Setup. Press ▲ or ▼ to skip Sensor and Calibration setup. When ALARM SETUP appears, press ON/OFF.
  4. Leave the alarm modes as they are. You’ll first see the vibrate and audible alarm options. Confirm each with ON/OFF without changing it.
  5. Go to the CO TWA screen. For each sensor you’ll step through LOW → HIGH → STEL → TWA. STEL and TWA only appear for toxic sensors (CO and H₂S). Press ON/OFF to accept each existing value until you reach TWA ALARM for CO.
  6. Set it to 20. Use ▲ / ▼ to change the value (holding a button moves it in steps of 10). Press ON/OFF to store it.
  7. Exit setup. Skip Time and Date, then confirm EXIT to return to measure mode.
  8. Check it. Turn the unit off and on again and watch the startup screens. The STEL and TWA setpoints are displayed during startup. Confirm CO TWA now reads 20.

Menus can vary slightly between firmware versions. Always follow the MSA operating manual for your specific model.

Tips from the field

  • Leave H₂S alone for now. Hydrogen sulphide is one of the chemicals still under national review, so its current limit continues until a decision is made. Don’t lower your H₂S setpoints yet. We’ll update this series when that changes.
  • Large fleet? MSA Link software or a GALAXY GX2 test station can push settings to multiple units. That’s faster and more consistent than changing them one by one.
  • Record the change. Note it in your equipment or asset register, and update any gas testing procedures, entry permits, SWMS and training material that still quote 30 ppm.
  • A TWA alarm isn’t permission to stay in. It warns about accumulated exposure over a shift. Your LOW and HIGH instantaneous alarms, together with your permit conditions, still decide whether people should be in that atmosphere at all.
  • Bump test before each day’s use, and keep calibration up to date.

Getting ready for WEL: your checklist

  • Identify every gas detector in your fleet with a CO sensor
  • Update the CO TWA alarm to 20 ppm before 1 December 2026
  • Confirm the change on each unit’s startup screen
  • Record the change in your asset register
  • Update procedures, permits, SWMS and training material
  • Brief workers and supervisors on the new limit